Easyencode 0731
Author: m | 2025-04-25
EasyEncode 0731 Full Version With License Key Latest.zip download from 4shared EasyEncode 0731 Full Version With License Key Latest.zip download from 4shared
EasyEncode 0731 Full Version With License Key Latest - 4shared
906-0731-01 Rev. J ZOLL Portable Critical Care Ventilator 2-17... Page 40 L. External Power Icon/Indicator -- Indicates the presence of external power. When no external power is sensed, the icon/indicator presents with a diagonal line. If an External Power Low or External Power Fail/Disconnect alarm occurs, the icon flashes off/on. 2-18 www.zoll.com 906-0731-01 Rev. J... Page 41 0, 50, and 100 cm H O. Above the bar graph, measured values for delivered minute volume, mean airway pressure (MAP), and actual breathing rate (set plus spontaneous) are displayed. 906-0731-01 Rev. J ZOLL Portable Critical Care Ventilator 2-19... Page 42: Pop Up Messages In addition, Pop Up messages can call for the operator to CONFIRM/SELECT that they are entering configurations where certain alarms are being suppressed, turned “off”, and/or cancelled. Pop Up Message Example 2-20 www.zoll.com 906-0731-01 Rev. J... Page 43 Cannot adjust low limit higher than high limit PIP SETTING CONFLICT Cannot adjust PIP target higher than PIP High Limit PIP SETTING CONFLICT Cannot adjust PIP target to less than 5 more than PEEP 906-0731-01 Rev. J ZOLL Portable Critical Care Ventilator 2-21... Page 44 Cannot adjust low limit higher than high limit LIMIT CONFLICT Cannot adjust low limit higher than V Setting LIMIT CONFLICT Cannot adjust V Setting above V High Limit LIMIT CONFLICT Cannot adjust V Setting above V Low Limit 2-22 www.zoll.com 906-0731-01 Rev. J... Page 45: Set-Up Warning! This unit’s USB connection does not provide any signal output or input to the operator. It is for the use of ZOLL-trained personnel only. The operator should not connect anything to the USB connection. 906-0731-01 Rev. J... Page 46: External Gas Sources RODUCT VERVIEW External Gas Sources The unit can operate using high-pressure oxygen from either oxygen cylinders or piped oxygen sources, as illustrated below: External Gas Sources 2-24 www.zoll.com 906-0731-01 Rev. J... Page 47: Fresh Gas/Emergency Air Intake The Fresh Gas/Emergency Air Intake also acts as an anti-asphyxia port in the event of a ventilator failure. 906-0731-01 Rev. J ZOLL Portable Critical Care Ventilator 2-25... Page 48: Ventilator Circuits: Connections Ventilator Circuit Device Connections Caution Always dispose of the circuit after single patient use following the institutional guidelines for biologically contaminated material. Reusing the circuit can result in cross contamination between patients. 2-26 www.zoll.com 906-0731-01 Rev. J... Page 49: Ventilator Circuits: Patient Connections Infant/Pediatric (Part #820-0107-00) and Pediatric/Adult (Part #820-0106-00). Similar circuits (differing only in hose length) are available for use in an MRI environment: Infant/Pediatric (Part #820-0131-00) and Pediatric/Adult (Part #820-0130-00). The circuits are illustrated below: Ventilator Circuit, Patient Connections 906-0731-01 Rev. J ZOLL Portable Critical Care Ventilator 2-27... Page 50: Ventilator Circuits: Resistance, Compliance, And Deadspace Resistance measured at Flow (l/min) ≥ Adult Pediatric 12-50 Infant
EasyEncode - Browse /EasyEncode v.1.0 (Alpha)/EasyEncode
Patient Movement, Check Sensor Placement, Check Sensor Operation **Turn Off Pulse Ox Monitoring** 4-10 www.zoll.com 906-0731-01 Rev. J... Page 89: Device Alarms 40 psig. Mitigation/Info: Manually Ventilate Patient, Connect 55 psig O2 then Restart, Check O2 Supply for Leaks,Replace Regulator **Replace/Service Ventilator** 906-0731-01 Rev. J ZOLL Portable Critical Care Ventilator 4-11... Page 90 1174 Off Set Self Check Failure The device is not able to calibrate the one or more transducers and is no longer able to operate safely. Mitigation/Info: Manually Ventilate Patient, Ventilator Not Functioning **Replace/Service Ventilator** 4-12 www.zoll.com 906-0731-01 Rev. J... Page 91 LCD is not visible. When this occurs, ventilation continues at the current settings or the backup mode settings and the High Priority alarm sounds. Mitigation/Info: Manually Ventilate Patient, Backup Ventilator Enabled **Replace/Service Ventilator** 906-0731-01 Rev. J ZOLL Portable Critical Care Ventilator 4-13... Page 92 SpO2 by adjusting the oxygen supply up or down to increase or decrease the amount of oxygen delivered to the patient. Mitigation/Info: Operation Switched to Compressor, Check O2 Supply Pressure, Check/Replace Regulator, Set FIO to 21%, Connect Low Flow O2, Monitor SpO2 **Replace/Service Ventilator** 4-14 www.zoll.com 906-0731-01 Rev. J... Page 93 (provided it is set properly as well). Mitigation/Info: Check Circuit for Kinked Hose/Tube, Check For Airway Obstruction, Suction Airway If Necessary, If Low Limit Set Correctly? **Manually Ventilate Patient** 906-0731-01 Rev. J ZOLL Portable Critical Care Ventilator 4-15... Page 94 Mitigation/Info: Internal Failure, SpO2/HR Not Available, Turn Off Pulse Ox **Replace/Service Ventilator** 2301 Internal Communication Failed The communication between the pulse oximeter and the unit fails. Mitigation/Info: Pulse Ox Module Failure, SpO2/HR Not Available, Turn Off Pulse Ox **Replace/Service Ventilator** 4-16 www.zoll.com 906-0731-01 Rev. J... Page 95 Communication between the compressor controller and SPM is lost, HP O is available to provide ventilation, and the operator has set the FIO to 100%. Mitigation/Info: Assure 55 psig O2, O2 Operation Only! **Replace/Service Ventilator** 906-0731-01 Rev. J ZOLL Portable Critical Care Ventilator 4-17... Page 96 Mitigation/Info: Barometric Pressure Sensor, Altitude Compensation Disable, Maintain Airway Pressure, Check Patient Chest Rise, Avoid Use At Varying Altitude **Replace/Service Ventilator** 4-18 www.zoll.com 906-0731-01 Rev. J... Page 97 The pulse oximeter does not recognize the connected sensor. The alarm can also occur when there is a broken sensor cable, inoperative LEDs, a fault is detected, and/or the sensor has failed. Mitigation/Info: Invalid Pulse Ox Sensor, Replace Sensor **Turn Off Pulse Ox Monitoring** 906-0731-01 Rev. J ZOLL Portable Critical Care Ventilator 4-19... Page 98 External Power Failed (Insufficient Current) The external power supply has insufficient current. When this occurs, the unit automatically switches to operation using the internal battery. Mitigation/Info: Internal Battery Operation, Check/Replace Power Supply, Change Power Source **Remove DC Connection** 4-20 www.zoll.com 906-0731-01 Rev. J... PageEasyEncode/ReadMe.txt at master jamalmazrui/EasyEncode
Contents Table of Contents Troubleshooting Bookmarks Portable Critical Care VentilatorOperator's GuideModels: EMV+, AEV, Eagle II731 Series906-0731-01 Rev. J Need help? Do you have a question about the 731 Series and is the answer not in the manual? Questions and answers Related Manuals for ZOLL 731 Series Summary of Contents for ZOLL 731 Series Page 1 Portable Critical Care Ventilator Operator’s Guide Models: EMV+, AEV, Eagle II 731 Series 906-0731-01 Rev. J... Page 2 The issue date for the ZOLL Portable Critical Care Ventilator Operator’s Guide (REF 906-0731-01 Rev. J) is June, 2017. Copyright © 2015 ZOLL Medical Corporation. All rights reserved. ZOLL, AEV, EMV+, Eagle II and Smart Help are registered trademarks or trademarks of ZOLL Medical Corporation in the United States and/or other countries. Page 3: Table Of Contents Software License ......................1-13 Limited Warranty ....................... 1-14 Service..........................1-15 Chapter 2 Product Overview ZOLL Ventilator Models ...................... 2-1 Common Features ...................... 2-1 EMV+ .......................... 2-2 AEV ..........................2-2 Eagle II ........................2-3 MRI Variants (Eagle II MRI and EMV+ MRI) .............. 2-3 Ventilator Controls and Indicators .................. Page 4 Use of Low Flow Oxygen ....................3-15 Use In an MRI Environment ....................3-18 Chapter 4 Alarms Smart Help Messages......................4-2 Alarm Priorities ......................4-3 Ventilator Alarm Categories ....................4-4 731 Series Ventilator Alarms....................4-9 Patient Alarms ......................4-9 Device Alarms ......................4-11 Environmental Alarms ....................4-22 www.zoll.com 906-0731-01 Rev. J... Page 5 Recharging Guidelines ..................... 5-10 Troubleshooting ........................ 5-11 Operator Correctable Problems ..................5-11 Operating Problems Requiring Service ................5-11 Storage ..........................5-11 Appendix A Specifications General ..........................A-1 Pulse Oximeter ........................A-3 IEC 60601-1........................A-4 Appendix B Accessories Index 906-0731-01 Rev. J ZOLL Portable Critical Care Ventilator... Page 6 Hazardous Environment Filters ....................3-11 Check Valve Connection To Breathing Circuit ................ 3-12 Oxygen Reservoir Bag Assembly ................... 3-17 ZOLL Ventilator Mounted On MRI Rolling Stand With IV Support Arm ........3-19 MR Labels on ZOLL Ventilator ....................3-20 Smart Help Example ......................... 4-1 Smart Help Display ........................ Page 7: General Information Operator’s Guide Updates An issue or revision date for this manual is shown on the front cover. If more than 3 years have elapsed since this date, contact ZOLL Medical Corporation to determine if additional product information updates are available. Page 8: Unpacking U.S.A. customers should call ZOLL Medical Corporation (1-973-882-1212). Customers outside of the U.S.A. should contact the nearest ZOLL authorized representative. If the shipping container is damaged, also notify the carrier. If there is no apparent sign of mechanical damage, read instructions contained within this manual before attempting operation. Page 9 External Power: Indicates the unit is operating using an external power source. No External Power: Indicates the unit is operating without an external power source. Internal Battery: Provides indication of battery capacity and charging. 906-0731-01 Rev. J ZOLL Portable Critical Care Ventilator... Page 10 Power Input. EasyEncode 0731 Full Version With License Key Latest.zip download from 4shared EasyEncode 0731 Full Version With License Key Latest.zip download from 4sharedEasyEncode/EasyEncode.txt at master jamalmazrui/EasyEncode
Arm for Rolling Cart (ferrous, not for MRI use) 800-0904-01 CCLAW (Critical Care Litter Attachment Widget) Mounting Bracket 703-0731-03 Carry-all Case with Foam Inserts, without AC Receptacle 703-0EMV-03 Carry-all Case with AC Receptacle 906-0731-01 Rev. J ZOLL Portable Critical Care Ventilator... Page 130 Mask, CPAP, #6, Large Adult with Harness (Case of 20) 712-0003-50 Mask, CPAP, #6, Large Adult with Harness (Case of 50) 334-0125-00 Harness, Mask, Universal 334-0125-00 Harness, Mask, Universal (Case of 10) 820-0132-00 600 ml ZOLL Test Lung, plastic/silicone www.zoll.com 906-0731-01 Rev. J... Page 131: Index Transducer (Patient Airway Pressure) .................. 2-7 USB Connector ........................2-7 Controls ..........................2-8–2-16 BPM (Breathing Rate) ....................2-11–2-12 Confirm/Select ........................2-13 FIO2 ............................ 2-10 HR ..........................2-9–2-10 Manual Breath ........................2-13 Menu ..........................2-13–2-16 Mode ..........................2-12–2-13 906-0731-01 Rev. J ZOLL Portable Critical Care Ventilator Index-1... Page 132 Rain and Snow ........................3-14 Hazardous Environment Filters ..................3-11–3-12 Bacterial/Viral ........................3-11 Change/Insertion ......................5-4–5-5 Check Valve on Breathing Circuit ..................3-12 Chemical/Biological ......................3-12 Foam Filter ..........................5-2 Fresh Gas/Emergency Air Intake Disk Filter ............... 5-3 Index-2 www.zoll.com 906-0731-01 Rev. J... Page 133 Principles ........................2-32–2-33 Specifications ........................A-3 Self-Check ..........................2-31 Service ..........................1-15–1-16 Set-up ............................2-23 Smart Help Messages ......................4-2–4-3 Software License ........................1-13 Specifications Accuracy/Precision ......................A-8 General ......................... A-1–A-2 906-0731-01 Rev. J ZOLL Portable Critical Care Ventilator Index-3... Page 134 Resistance, Compliance, and Deadspace ..............2-28–2-29 Ventilator Models AEV ............................2-2 Common Features ......................... 2-1 Eagle II ..........................2-3 EMV+ ........................... 2-2 MRI Variants ........................2-3 Warnings ..........................1-8–1-11 General ........................... 1-8–1-9 Pulse Oximeter ......................1-10–1-11 Index-4 www.zoll.com 906-0731-01 Rev. J...EasyEncode/EasyEncode.htm at master jamalmazrui/EasyEncode
Violate (PEEP + PS) 3 cm H Note: Negative pressure is not available in the expiratory phase. 906-0731-01 Rev. J ZOLL Portable Critical Care Ventilator... Page 60: Additional Adjuncts Of Operation Insufficient Flow (Alarm #2095) ® ™ To start NPPV in either EMV+ and/or Eagle II 1. Press MODE button to highlight mode. Select CPAP and press Accept. Press MODE button until the PPV field is selected. Adjust to NPPV. www.zoll.com 906-0731-01 Rev. J... Page 61 NPPV into CPAP Default Value NPPV to P Default Value High BPM Alarm Low BPM Alarm Low Airway Pressure Alarm (PIP Target = -3) (PIP Target = -3) PEEP No change No change High Limit 906-0731-01 Rev. J ZOLL Portable Critical Care Ventilator... Page 62: Spontaneous/Assisted Breath Trigger PATIENT DISCONNECT alarm should activate. Partially close the patient port to reset the PATIENT DISCONNECT alarm. With no other alarms occurring, remove external power from the ventilator. The EXTERNAL POWER LOW/DISCONNECT alarms should activate. Reconnect external power to reset alarms. www.zoll.com 906-0731-01 Rev. J... Page 63: To Begin Ventilating 2. Other airways can include: laryngeal mask airway, esophageal obturator airway, and combination esophageal-tracheal tubes though their use should be under the direction of the attending physician. See AARC Clinical Practice Guideline, Management of Airway Emergencies Respiratory Care 1995;40(7): 749-760 Section 10.2.2. 906-0731-01 Rev. J ZOLL Portable Critical Care Ventilator... Page 64: To Change Settings Press the CONFIRM/SELECT button switch “√” to complete the value change. 3. If the accuracy of any measurements does not seem reasonable, first check the patient’s vital signs by alternate means and then check the pulse oximeter for proper functioning. www.zoll.com 906-0731-01 Rev. J... Page 65: Back Up Ventilator HMEs or HMEFs attach between the disposable ventilator circuit and patient’s endotracheal tube. Be sure to follow all instructions provided by the manufacturer. 906-0731-01 Rev. J ZOLL Portable Critical Care Ventilator... Page 66 4 (infant/ ped) 10 ml ZOLL does not offer a heated humidifier option for the unit. Users are cautioned to carefully consider the ramifications of such use and the effect it may have upon device performance and the patient’s comfort. Such humidifiers have been shown to increase the work of breathing in portable ventilators . Page 67: Hazardous Environment Filters 22 mm conical fitting into the Fresh Gas/Emergency Air Intake. Caution If filters have been exposed to biological matter, dispose of them following the Universal Precaution procedures for your facility. 906-0731-01 Rev. J ZOLL Portable Critical Care Ventilator 3-11... Page 68: Chemical/Biological Filter Use Use of the 3M canister does not constitute endorsement or recommendation of the 3M device. Use and selection of the appropriate filter should always be under the direction of the Incident Commander. 3-12 www.zoll.com 906-0731-01 Rev.EasyEncode - Browse /EasyEncode v.1.0 (Alpha) at
Matter, dispose of them following Universal Precaution procedures for your facility. Note: Do not attempt to clean this filter and do not operate internal compressor without a filter in place. 906-0731-01 Rev. J ZOLL Portable Critical Care Ventilator... Page 110: Internal Filter Change/Insertion Compressor Inlet Fitting Assembly to the SPM Chassis. Lift the two (2) segments of the Compressor Inlet Fitting Assembly away from the unit. If the two segments come apart, do not lose the gasket that seats between the parts. Alignment Pins www.zoll.com 906-0731-01 Rev. J... Page 111 (4) 8-32 x 3 Phillips Flat Head screws. Momentarily turn the unit’s POWER switch to its “ON” position to confirm operating power. A DISCONNECT alarm will sound. Turn the unit’s POWER switch to its “OFF” position. 906-0731-01 Rev. J ZOLL Portable Critical Care Ventilator... Page 112: Post-Contaminated Environment Cleaning ZOLL suggests that each facility have in place a procedure for the cleaning and disinfection of its medical equipment and that these procedures to consulted for further guidance. Page 113: Battery Maintenance 2 hours. Following these simple guidelines will prevent premature charge depletion and reduction of battery life. Caution Only use the Power Supply provided with the unit (ZOLL PN 703-0731-01). Use of any other power supply could cause damage or create a fire and/or destroy the battery and unit. Caution... Page 114 “disappears” as battery capacity is reduced by a 5% increment. The Battery icon will flash off/on when a BATTERY POWER LOW alarm occurs. The icon will flash off/on and present a diagonal line when no battery is connected. www.zoll.com 906-0731-01 Rev. J... Page 115: Internal Battery Change/Insertion Phillips Head Screwdriver Warning! Before attempting to replace the battery pack (ZOLL PN 703-0731-01), make sure that external power is disconnected and the unit’s Power Switch is set to “OFF”. Procedure: Remove the four (4) 6-32 x 5/16 Phillips Pan Head screws located at the bottom of the unit. Page 116: Recharging Guidelines F even though the unit is still connected to external power. ° ° Under these conditions, battery temperature can get as high as 10 C/50 F above the ambient temperature. Charging will automatically start when the ambient temperature drops. ZOLL Ventilator Storage Case 5-10 www.zoll.com 906-0731-01 Rev. J... Page 117: Troubleshooting Troubleshooting Authorization to service this instrument by other than factory-trained and certified personnel will not be given, nor does ZOLL assume any responsibility and/or liability resulting from such unauthorized servicing. Operator Correctable Problems Common problems may be quickly rectified by the operator following the alarm mitigation instructions. Page 118 Following 6 months of continuous storage/non-use, or longer, the device should be examined, operationally tested, and its batteries recharged before patient use is attempted. Service may be required.. EasyEncode 0731 Full Version With License Key Latest.zip download from 4sharedComments
906-0731-01 Rev. J ZOLL Portable Critical Care Ventilator 2-17... Page 40 L. External Power Icon/Indicator -- Indicates the presence of external power. When no external power is sensed, the icon/indicator presents with a diagonal line. If an External Power Low or External Power Fail/Disconnect alarm occurs, the icon flashes off/on. 2-18 www.zoll.com 906-0731-01 Rev. J... Page 41 0, 50, and 100 cm H O. Above the bar graph, measured values for delivered minute volume, mean airway pressure (MAP), and actual breathing rate (set plus spontaneous) are displayed. 906-0731-01 Rev. J ZOLL Portable Critical Care Ventilator 2-19... Page 42: Pop Up Messages In addition, Pop Up messages can call for the operator to CONFIRM/SELECT that they are entering configurations where certain alarms are being suppressed, turned “off”, and/or cancelled. Pop Up Message Example 2-20 www.zoll.com 906-0731-01 Rev. J... Page 43 Cannot adjust low limit higher than high limit PIP SETTING CONFLICT Cannot adjust PIP target higher than PIP High Limit PIP SETTING CONFLICT Cannot adjust PIP target to less than 5 more than PEEP 906-0731-01 Rev. J ZOLL Portable Critical Care Ventilator 2-21... Page 44 Cannot adjust low limit higher than high limit LIMIT CONFLICT Cannot adjust low limit higher than V Setting LIMIT CONFLICT Cannot adjust V Setting above V High Limit LIMIT CONFLICT Cannot adjust V Setting above V Low Limit 2-22 www.zoll.com 906-0731-01 Rev. J... Page 45: Set-Up Warning! This unit’s USB connection does not provide any signal output or input to the operator. It is for the use of ZOLL-trained personnel only. The operator should not connect anything to the USB connection. 906-0731-01 Rev. J... Page 46: External Gas Sources RODUCT VERVIEW External Gas Sources The unit can operate using high-pressure oxygen from either oxygen cylinders or piped oxygen sources, as illustrated below: External Gas Sources 2-24 www.zoll.com 906-0731-01 Rev. J... Page 47: Fresh Gas/Emergency Air Intake The Fresh Gas/Emergency Air Intake also acts as an anti-asphyxia port in the event of a ventilator failure. 906-0731-01 Rev. J ZOLL Portable Critical Care Ventilator 2-25... Page 48: Ventilator Circuits: Connections Ventilator Circuit Device Connections Caution Always dispose of the circuit after single patient use following the institutional guidelines for biologically contaminated material. Reusing the circuit can result in cross contamination between patients. 2-26 www.zoll.com 906-0731-01 Rev. J... Page 49: Ventilator Circuits: Patient Connections Infant/Pediatric (Part #820-0107-00) and Pediatric/Adult (Part #820-0106-00). Similar circuits (differing only in hose length) are available for use in an MRI environment: Infant/Pediatric (Part #820-0131-00) and Pediatric/Adult (Part #820-0130-00). The circuits are illustrated below: Ventilator Circuit, Patient Connections 906-0731-01 Rev. J ZOLL Portable Critical Care Ventilator 2-27... Page 50: Ventilator Circuits: Resistance, Compliance, And Deadspace Resistance measured at Flow (l/min) ≥ Adult Pediatric 12-50 Infant
2025-04-03Patient Movement, Check Sensor Placement, Check Sensor Operation **Turn Off Pulse Ox Monitoring** 4-10 www.zoll.com 906-0731-01 Rev. J... Page 89: Device Alarms 40 psig. Mitigation/Info: Manually Ventilate Patient, Connect 55 psig O2 then Restart, Check O2 Supply for Leaks,Replace Regulator **Replace/Service Ventilator** 906-0731-01 Rev. J ZOLL Portable Critical Care Ventilator 4-11... Page 90 1174 Off Set Self Check Failure The device is not able to calibrate the one or more transducers and is no longer able to operate safely. Mitigation/Info: Manually Ventilate Patient, Ventilator Not Functioning **Replace/Service Ventilator** 4-12 www.zoll.com 906-0731-01 Rev. J... Page 91 LCD is not visible. When this occurs, ventilation continues at the current settings or the backup mode settings and the High Priority alarm sounds. Mitigation/Info: Manually Ventilate Patient, Backup Ventilator Enabled **Replace/Service Ventilator** 906-0731-01 Rev. J ZOLL Portable Critical Care Ventilator 4-13... Page 92 SpO2 by adjusting the oxygen supply up or down to increase or decrease the amount of oxygen delivered to the patient. Mitigation/Info: Operation Switched to Compressor, Check O2 Supply Pressure, Check/Replace Regulator, Set FIO to 21%, Connect Low Flow O2, Monitor SpO2 **Replace/Service Ventilator** 4-14 www.zoll.com 906-0731-01 Rev. J... Page 93 (provided it is set properly as well). Mitigation/Info: Check Circuit for Kinked Hose/Tube, Check For Airway Obstruction, Suction Airway If Necessary, If Low Limit Set Correctly? **Manually Ventilate Patient** 906-0731-01 Rev. J ZOLL Portable Critical Care Ventilator 4-15... Page 94 Mitigation/Info: Internal Failure, SpO2/HR Not Available, Turn Off Pulse Ox **Replace/Service Ventilator** 2301 Internal Communication Failed The communication between the pulse oximeter and the unit fails. Mitigation/Info: Pulse Ox Module Failure, SpO2/HR Not Available, Turn Off Pulse Ox **Replace/Service Ventilator** 4-16 www.zoll.com 906-0731-01 Rev. J... Page 95 Communication between the compressor controller and SPM is lost, HP O is available to provide ventilation, and the operator has set the FIO to 100%. Mitigation/Info: Assure 55 psig O2, O2 Operation Only! **Replace/Service Ventilator** 906-0731-01 Rev. J ZOLL Portable Critical Care Ventilator 4-17... Page 96 Mitigation/Info: Barometric Pressure Sensor, Altitude Compensation Disable, Maintain Airway Pressure, Check Patient Chest Rise, Avoid Use At Varying Altitude **Replace/Service Ventilator** 4-18 www.zoll.com 906-0731-01 Rev. J... Page 97 The pulse oximeter does not recognize the connected sensor. The alarm can also occur when there is a broken sensor cable, inoperative LEDs, a fault is detected, and/or the sensor has failed. Mitigation/Info: Invalid Pulse Ox Sensor, Replace Sensor **Turn Off Pulse Ox Monitoring** 906-0731-01 Rev. J ZOLL Portable Critical Care Ventilator 4-19... Page 98 External Power Failed (Insufficient Current) The external power supply has insufficient current. When this occurs, the unit automatically switches to operation using the internal battery. Mitigation/Info: Internal Battery Operation, Check/Replace Power Supply, Change Power Source **Remove DC Connection** 4-20 www.zoll.com 906-0731-01 Rev. J... Page
2025-04-07Arm for Rolling Cart (ferrous, not for MRI use) 800-0904-01 CCLAW (Critical Care Litter Attachment Widget) Mounting Bracket 703-0731-03 Carry-all Case with Foam Inserts, without AC Receptacle 703-0EMV-03 Carry-all Case with AC Receptacle 906-0731-01 Rev. J ZOLL Portable Critical Care Ventilator... Page 130 Mask, CPAP, #6, Large Adult with Harness (Case of 20) 712-0003-50 Mask, CPAP, #6, Large Adult with Harness (Case of 50) 334-0125-00 Harness, Mask, Universal 334-0125-00 Harness, Mask, Universal (Case of 10) 820-0132-00 600 ml ZOLL Test Lung, plastic/silicone www.zoll.com 906-0731-01 Rev. J... Page 131: Index Transducer (Patient Airway Pressure) .................. 2-7 USB Connector ........................2-7 Controls ..........................2-8–2-16 BPM (Breathing Rate) ....................2-11–2-12 Confirm/Select ........................2-13 FIO2 ............................ 2-10 HR ..........................2-9–2-10 Manual Breath ........................2-13 Menu ..........................2-13–2-16 Mode ..........................2-12–2-13 906-0731-01 Rev. J ZOLL Portable Critical Care Ventilator Index-1... Page 132 Rain and Snow ........................3-14 Hazardous Environment Filters ..................3-11–3-12 Bacterial/Viral ........................3-11 Change/Insertion ......................5-4–5-5 Check Valve on Breathing Circuit ..................3-12 Chemical/Biological ......................3-12 Foam Filter ..........................5-2 Fresh Gas/Emergency Air Intake Disk Filter ............... 5-3 Index-2 www.zoll.com 906-0731-01 Rev. J... Page 133 Principles ........................2-32–2-33 Specifications ........................A-3 Self-Check ..........................2-31 Service ..........................1-15–1-16 Set-up ............................2-23 Smart Help Messages ......................4-2–4-3 Software License ........................1-13 Specifications Accuracy/Precision ......................A-8 General ......................... A-1–A-2 906-0731-01 Rev. J ZOLL Portable Critical Care Ventilator Index-3... Page 134 Resistance, Compliance, and Deadspace ..............2-28–2-29 Ventilator Models AEV ............................2-2 Common Features ......................... 2-1 Eagle II ..........................2-3 EMV+ ........................... 2-2 MRI Variants ........................2-3 Warnings ..........................1-8–1-11 General ........................... 1-8–1-9 Pulse Oximeter ......................1-10–1-11 Index-4 www.zoll.com 906-0731-01 Rev. J...
2025-04-19Violate (PEEP + PS) 3 cm H Note: Negative pressure is not available in the expiratory phase. 906-0731-01 Rev. J ZOLL Portable Critical Care Ventilator... Page 60: Additional Adjuncts Of Operation Insufficient Flow (Alarm #2095) ® ™ To start NPPV in either EMV+ and/or Eagle II 1. Press MODE button to highlight mode. Select CPAP and press Accept. Press MODE button until the PPV field is selected. Adjust to NPPV. www.zoll.com 906-0731-01 Rev. J... Page 61 NPPV into CPAP Default Value NPPV to P Default Value High BPM Alarm Low BPM Alarm Low Airway Pressure Alarm (PIP Target = -3) (PIP Target = -3) PEEP No change No change High Limit 906-0731-01 Rev. J ZOLL Portable Critical Care Ventilator... Page 62: Spontaneous/Assisted Breath Trigger PATIENT DISCONNECT alarm should activate. Partially close the patient port to reset the PATIENT DISCONNECT alarm. With no other alarms occurring, remove external power from the ventilator. The EXTERNAL POWER LOW/DISCONNECT alarms should activate. Reconnect external power to reset alarms. www.zoll.com 906-0731-01 Rev. J... Page 63: To Begin Ventilating 2. Other airways can include: laryngeal mask airway, esophageal obturator airway, and combination esophageal-tracheal tubes though their use should be under the direction of the attending physician. See AARC Clinical Practice Guideline, Management of Airway Emergencies Respiratory Care 1995;40(7): 749-760 Section 10.2.2. 906-0731-01 Rev. J ZOLL Portable Critical Care Ventilator... Page 64: To Change Settings Press the CONFIRM/SELECT button switch “√” to complete the value change. 3. If the accuracy of any measurements does not seem reasonable, first check the patient’s vital signs by alternate means and then check the pulse oximeter for proper functioning. www.zoll.com 906-0731-01 Rev. J... Page 65: Back Up Ventilator HMEs or HMEFs attach between the disposable ventilator circuit and patient’s endotracheal tube. Be sure to follow all instructions provided by the manufacturer. 906-0731-01 Rev. J ZOLL Portable Critical Care Ventilator... Page 66 4 (infant/ ped) 10 ml ZOLL does not offer a heated humidifier option for the unit. Users are cautioned to carefully consider the ramifications of such use and the effect it may have upon device performance and the patient’s comfort. Such humidifiers have been shown to increase the work of breathing in portable ventilators . Page 67: Hazardous Environment Filters 22 mm conical fitting into the Fresh Gas/Emergency Air Intake. Caution If filters have been exposed to biological matter, dispose of them following the Universal Precaution procedures for your facility. 906-0731-01 Rev. J ZOLL Portable Critical Care Ventilator 3-11... Page 68: Chemical/Biological Filter Use Use of the 3M canister does not constitute endorsement or recommendation of the 3M device. Use and selection of the appropriate filter should always be under the direction of the Incident Commander. 3-12 www.zoll.com 906-0731-01 Rev.
2025-04-23Longer breathing circuits may increase the risk of self-triggering of ventilator breaths. Reducing the pressure trigger sensitivity may solve this problem. 3-22 www.zoll.com 906-0731-01 Rev. J... Page 79: Chapter 4 Alarms Chapter 4 Alarms ™ The ZOLL Portable Critical Care Ventilator uses Smart Help messages that provide a comprehensive suite of alarms to alert operators and guide their actions to resolve alarm conditions and assure patient safety. At the onset of an alarm, the screen displays the alarm name and then a series of context-sensitive Smart Help messages. Page 80: Smart Help Messages Alarms are prioritized based on the risk to the patient. The alarm with the greatest risk to the patient is always presented first. All messages are context-based and suggest what is causing the condition and/or how it can be resolved. Smart Help Display www.zoll.com 906-0731-01 Rev. J... Page 81: Alarm Priorities Pressing the MUTE/CANCEL button cancels the audible signal. If the alarm is not resolved, the yellow LED remains illuminated to remind the operator of the fault or failure. Some Low Priority alarms may be canceled to avoid nuisance alarms. 906-0731-01 Rev. J ZOLL Portable Critical Care Ventilator... Page 82: Ventilator Alarm Categories >80 psig, the unit automatically shuts down to prevent harm to the patient and/or ≥ ≤ damage to the unit. When the pressure is 75 and 80 psig, a Low Priority alarm sounds, warning the operator of the potential shutdown should pressure increase beyond the current level. www.zoll.com 906-0731-01 Rev. J... Page 83 Resolution of this fault is typically associated with problems causing restrictions to the exhalation valve. Pressing the MUTE/CANCEL button mutes the audible alarm for 30 seconds. 906-0731-01 Rev. J ZOLL Portable Critical Care Ventilator... Page 84 Pressing the MUTE/CANCEL button mutes the audible alarm, but the persistent message and yellow LED remain. www.zoll.com 906-0731-01 Rev. J... Page 85 Pressing the MUTE/CANCEL button cancels the alarm. To clear the alarm, reposition the sensor to a better site or suspend SpO monitoring by placing it in standby (stby). 906-0731-01 Rev. J ZOLL Portable Critical Care Ventilator... Page 86 Pressing the MUTE/CANCEL button cancels the alarm. To clear the alarm, use an appropriate sensor or suspend SpO monitoring by placing it in standby (stby). www.zoll.com 906-0731-01 Rev. J... Page 87: 731 Series Ventilator Alarms 731 Series Ventilator Alarms The ZOLL 731 Series Ventilator displays alarms that indicate the following types of alarm conditions: Patient Alarms -- triggered by the actions or condition of the patient • Device Alarms -- triggered by a malfunction or internal fault with the device •... Page 88 The pulse oximeter determines the quality of the input signal is low due to excessive motion or artifact. Mitigation/Info: Signal Artifact, Minimize
2025-04-1599 The unit software detects that it has not been calibrated with the SPM that is inside the unit. This fault occurs when the biomedical technician fails to recalibrate the unit following an SPM change or service. Mitigation/Info: Hardware Compatibility Failure, Update Calibration Records **Replace/Service Ventilator** 906-0731-01 Rev. J ZOLL Portable Critical Care Ventilator 4-21... Page 100: Environmental Alarms The ambient pressure transducer detects an altitude >25,000 feet (7620 meters). Beyond this altitude, compensation remains fixed at the 25,000 ft compensation level. Mitigation/Info: Excessive Altitude Detected, Beyond Altitude Compensation Limit, Maintain Airway Pressure, Check Patient Chest Rise, Monitor Ventilator/Patient **Reduce Altitude** 4-22 www.zoll.com 906-0731-01 Rev. J... Page 101 Mitigation/Info: Battery Too Hot to Discharge, DO NOT Remove External Power!, Remove Padded Case, Assure Ability to Manually Ventilate Patient, Shade Patient and Ventilator **Move to Cooler Location** 906-0731-01 Rev. J ZOLL Portable Critical Care Ventilator 4-23... Page 102 This failure may be caused by a significant blockage of the exhalation valve or an occlusion/kink in the exhalation valve tube. Mitigation/Info: Manually Ventilate Patient, Replace Circuit and Restart **Replace/Service Ventilator** 4-24 www.zoll.com 906-0731-01 Rev. J... Page 103: Alarm Turn Off And Cancellation Spontaneous Breath-PIP Low 2172 Spontaneous Breath-V High 2173 Spontaneous Breath-V 2300 Pulse Ox Module Failed 2301 Internal Communication Failed 2314 SpO2 Sensor Off Patient 2401 SpO2 Low 2410 Heart Rate High 906-0731-01 Rev. J ZOLL Portable Critical Care Ventilator 4-25... Page 104: Alarm Turn Off At Extreme Range Limits High Breath High Heart Low Heart High PIP Low Low SpO2 Rate Rate Rate 99 BPM 3 cm H 2000 ml 0 ml 240 BPM 30 BPM 4-26 www.zoll.com 906-0731-01 Rev. J... Page 105: Alarm Cancellation In Alarm Configuration Menu Self Check Fault, calibration due (Alarm #3120) RTC Battery Fault (Battery Low) (Alarm #3110) Incomplete Exhalation (Alarm #3110) PEEP Leak (Alarm #2090) Fresh Gas Intake Fault (Alarm #3031) Patient Inspiratory Demand Not Met (Alarm #3092) 906-0731-01 Rev. J ZOLL Portable Critical Care Ventilator 4-27... Page 106 LARMS 4-28 www.zoll.com 906-0731-01 Rev. J... Page 107: Chapter 5 Maintenance Chapter 5 Maintenance The ZOLL ventilator must be maintained to be ready for immediate use. The device should be incorporated into a regular preventative maintenance program to insure compliance with operating specifications. Calibration Checks Calibration checks should be done every 12 months or 1500 service hours, unless significant usage warrants a shorter period between preventative maintenance inspections. Page 108: General Cleaning To insure compliance with operating specifications, it is the user’s responsibility to insure that periodic preventative maintenance is performed. It is recommended that the In-Field Calibration Check be performed by ZOLL or a certified ZOLL service facility. Page 109: Fresh Gas/Emergency Air Intake Disk Filter Replacement If filters have been exposed to biological
2025-04-24